RSM logo
Annals of Clinical Biochemistry

Home Current issue Browse archive Alerts About the journal Feedback
 
Ann Clin Biochem 2009;46:79-81
doi:10.1258/acb.2008.008135
© 2009 Association for Clinical Biochemistry

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow An erratum has been published
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (3)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Carter, G. D
Right arrow Articles by Jones, J. C
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Short Report

Use of a common standard improves the performance of liquid chromatography-tandem mass spectrometry methods for serum 25-hydroxyvitamin-D

Graham D Carter and Julia C Jones


Clinical Chemistry Department, Imperial College Healthcare NHS Trust, Charing Cross Hospital, London W6 8RF, UK


Corresponding author: Julia C Jones. Email: Julia.jones{at}imperial.ac.uk


Background: Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is becoming increasingly popular for measuring 25-hydroxyvitamin-D (25-OH-D). Results submitted to the International Quality Assessment Scheme (DEQAS) have shown poor interlaboratory agreement. We investigated whether the use of a common standard would reduce interlaboratory imprecision.

Methods: A commercial standard and two controls were distributed with the DEQAS samples in January 2008. Participants were asked to calculate the results of samples and controls using their usual standard and the commercial standard. A method questionnaire was also distributed.

Results: Use of a common standard reduced the mean interlaboratory imprecision (coefficient of variation [CV]) for total 25-OH-D from 16.4% (in-house standards) to 10.4% (common standard). For 25-OH-D3 and 25-OH-D2, the mean CVs were reduced from 16.7% and 21.1% to 8.5% and 12.6%, respectively. Mean values obtained for total 25-OH-D using the common standard were higher by 6.1%.

Conclusions: Use of a common standard improved agreement among laboratories using LC-MS/MS methods for 25-OH-D. This suggests that problems with assay standardization contribute to interlaboratory imprecision. This may be related to the nature of the matrix used for working standards or errors in the calibration of stock standard solutions of 25-OH-D. Some participants used a gravimetric method, others UV spectrophotometry, to establish the concentration of stock solutions. Among the latter group there was uncertainty over the molar absorption coefficient of 25-OH-D solutions. We conclude that LC-MS/MS is not yet sufficiently robust to become the reference method for 25-OH-D and that gas chromatography-mass spectrometry might be a more suitable candidate.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
Clin. Chem.Home page
G. D. Carter
25-Hydroxyvitamin D Assays: The Quest For Accuracy
Clin. Chem., July 1, 2009; 55(7): 1300 - 1302.
[Full Text] [PDF]


Home page
Ann Clin BiochemHome page
W. D Fraser
Standardization of vitamin D assays: art or science?
Ann Clin Biochem, January 1, 2009; 46(1): 3 - 4.
[Full Text] [PDF]



MDU Exam Doctor