RSM logo
Annals of Clinical Biochemistry

Home Current issue Browse archive Alerts About the journal Feedback
 
First published on 16 October 2009
Ann Clin Biochem
doi:acb.2009.008191
© 2009 Association for Clinical Biochemistry

This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Griffith, C A
Right arrow Articles by Keevil, B G
PubMed
Right arrow PubMed Citation
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Original Article

Development of a method to measure plasma and whole blood choline by liquid chromatography tandem mass spectrometry

C A Griffith1, L J Owen1, R Body2, G McDowell2 and B G Keevil1


1 University Hospital of South Manchester, Southmoor Road, Wythenshawe, Manchester M23 9LT; 2 Manchester Royal Infirmary, Oxford Road, Manchester M13 9WL, UK


Corresponding author: C A Griffith, Department of Clinical Biochemistry, University Hospital of South Manchester, Southmoor Road, Wythenshawe, Manchester M23 9LT, UK. Email: caroline.griffith{at}nhs.net


Background: Current gold standard markers for myocardial damage are troponins I and T, which are both sensitive and specific for the detection of myocardial infarction, but require up to 6 h to become reliably elevated in serum. Investigation into markers with potential to identify patients with early ischaemic changes is therefore intense. Choline is reported to be prognostic in patients presenting with acute coronary syndromes via its release from ischaemic cell membranes.

Methods: Liquid chromatography tandem mass spectrometry was used to develop a method to quantitate choline in plasma and blood. The method involves addition of a deuterated internal standard to an aliquot of plasma or blood followed by organic solvent addition, which precipitates the proteins in the sample. Preparation was carried out directly into a 96-deep-well plate. Chromatography of choline used a strong cation exchange column and separation used a Waters Atlantis dC18 analytical column positioned directly before the mass spectrometer source, allowing on-line preanalytical clean up of the sample.

Results: The lower limit of quantitation was 0.38 µmol/L, linearity was observed up to 754 µmol/L, with a working concentration range of 0.38–224 µmol/L, inter- and intra-assay coefficients of variation were <6% and <4%, respectively. Samples were stable throughout five freeze–thaw cycles and recovery was between 94% and 114%.

Conclusions: The assay was successfully validated in accordance with FDA guidelines and is suitable for quantitation of choline in research and clinical settings.