Can serum free light chains replace urine electrophoresis in the detection of monoclonal gammopathies?

Ann Clin Biochem 2007;44:516-522
doi:10.1258/000456307782268084
© 2007 Association for Clinical Biochemistry

 

This Article
Right arrow

Full Text (PDF)

Right arrow
Alert me when this article is cited
Right arrow
Alert me if a correction is posted
Services
Right arrow
Email this article to a friend
Right arrow

Similar articles in this journal

Right arrow
Similar articles in PubMed
Right arrow
Alert me to new issues of the journal
Right arrow
Download to citation manager
Right arrow
Citing Articles
Right arrow
Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow
Articles by Beetham, R.
Right arrow
Articles by James, J. A
Right arrow Search for Related Content
PubMed
Right arrow
PubMed Citation
Social Bookmarking

What’s this?

Original Articles


Robert Beetham,
Julie Wassell,
Michael J Wallage,
Alastair J Whiteway and
Jacky A James


Department of Clinical Haematology, North Bristol NHS, Trust, Frenchay Hospital Bristol BS16 1LE, UK


Background: We investigated how sensitive serum free light chain (FLC) analysis was for the detection of Bence-Jones protein (BJP) and whether a serum {kappa}/{lambda} ratio could replace urine electrophoresisas part of the investigation algorithm for monoclonal gammopathy.

Methods: Serum {kappa} and {lambda} FLC analysis was performed on 932 consecutive patients investigated for monoclonal gammopathy, along with serum electrophoresis, serum immunofixation where appropriate, and in 483 individuals urine immunofixation. A reference interval for the {kappa}/{lambda} ratio was derived from all patients who had normal serum and urine electrophoresis and no B-cell dyscrasia (n =312).

Results: The 100% reference interval was 0.21-1.44 (median 0.50), lower than that previously published. From the 483 patients who provided urines, BJP was detected in 34, with an abnormal {kappa}/{lambda} ratio in 26 of these. Using the reference interval, sensitivity, specificity, negative and positive predictive values of an abnormal {kappa}/{lambda} ratio for the presence of BJP were 0.76, 0.96, 0.98 and 0.59 respectively. Seven patients had an abnormal {kappa}/{lambda} ratio with noparaprotein.

Conclusion: All {kappa}/{lambda} results should be interpreted against a reference interval appropriate for the analyser in use and the population under study. Some urines will be positive for BJP when the serum {kappa}/{lambda} ratio is normal, but there is little difference in outcomewhether urine for BJP or serum FLC analysis is included in thescreening strategy for monoclonal gammopathies. On balance,we believe that it is not yet time to dispense with urine BJPin the investigation of patients with suspected monoclonal gammopathy.


CiteULike    Complore    Connotea    Del.icio.us    Digg    Reddit    Technorati    What’s this?






This article has been cited by other articles:


J. M. Abadie, K.H. van Hoeven, and J. M. Wells
Are Renal Reference Intervals Required When Screening for Plasma Cell Disorders With Serum Free Light Chains and Serum Protein Electrophoresis?
Am J Clin Pathol,

February 1, 2009;
131(2):
166 – 171.

[Abstract]
[Full Text]
[PDF]


Home page Ann Clin BiochemHome page

G P Mead, H D Carr-Smith, and A R Bradwell
Free light chains
Ann Clin Biochem,

July 1, 2008;
45(4):
444 – 444.

[Full Text]
[PDF]